BNC-2110 Installation Guide 4 ni.com
1. Place the BNC-2110 near the host computer or PXI/PXI Express chassis or use the optional DIN
Rail Mounting kit for UMI-FLEX-6 and BNC boxes (part number 777972-01), which you can
order from National Instruments at ni.com.
Caution Do not connect input voltages greater than 42.4 V
pk
/60 VDC to the BNC-2110. The
BNC-2110 is not designed for any input voltages greater than 42.4 V
pk
/60 VDC, even if a
user-installed voltage divider reduces the voltage to within the input range of the DAQ device. Input
voltages greater than 42.4 V
pk
/60 VDC can damage the BNC-2110, all devices connected to it, and
the host computer. Overvoltage can also cause an electric shock hazard for the operator. National
Instruments is not liable for damage or injury resulting from such misuse.
2. Connect the BNC-2110 to the DAQ device using the appropriate cable for your DAQ device, as
listed in Table 1.
The power indicator light, shown in Figure 1, lights. If it does not light, check the cable and wiring
connections.
3. Launch Measurement & Automation Explorer (MAX), confirm that your DAQ device is
recognized, and configure your device settings. Refer to the DAQ Getting Started Guide for more
information.
4. Connect signals to the BNC connectors and spring terminal block as described in the following
sections.
Note With NI-DAQmx, National Instruments has revised its terminal names so they are easier to
understand and more consistent among NI hardware and software products. The revised terminal
names used in this document are usually similar to the names they replace. For a complete list of
Traditional NI-DAQ (Legacy) terminal names and their NI-DAQmx equivalents, refer to the
Terminal Name Equivalents table in the NI-DAQmx Help.
5. Test specific device functionality, such as the ability to send and receive data with MAX test panels.
Refer to the DAQ Getting Started Guide for more detailed information about running test panels
in MAX.
Table 1. BNC-2110 Cabling Options
Number of Pins DAQ Device Recommended Cable(s)
68-pin NI 6715 AO Series
DAQCard E Series
NI PCI/PCIe/PXI/PXIe M Series
*
NI 6143 S Series
SHC68-68-EPM or RC68-68
NI 672x AO Series
SH68-C68-S
NI 6711/673x AO Series
PCI/PXI E Series
USB Mass Termination M Series
*
NI 6110/6111/612x/613x S Series
SH68-68-EP or R6868
NI 6713 AO Series
NI 6115/6120 S Series
SH68-68-EP
100-pin PCI/PXI E Series SH1006868
*
You cannot connect the BNC-2110 to Connector 1 of NI 6225/6255 devices.
You cannot connect the BNC-2110 to the AO 8–31 Connector of NI 6723 devices.